High-frequency, dielectric spectroscopy for the detection of electrophysiological/biophysical differences in different bacteria types and concentrations
Date
2018Author
Russel, MohammadSophocleous, Marios
JiaJia, Shan
Xu, Weiping
Xiao, Lehui
Maskow, Thomas
Alam, Mahbub
Georgiou, Julius
ISSN
0003-2670Source
Analytica Chimica ActaVolume
1028Pages
86-95Google Scholar check
Metadata
Show full item recordAbstract
This paper describes a novel technique to quantify and identify bacterial cultures of Bacillus Subtilis (2.10–1.30 × 109 CFU mL−1) and Escherichia Coli (1.60–1.00 × 109 CFU mL−1), in corn oil using dielectric spectroscopy at elevated frequencies of 0.0100–20.0 GHz. This technique is using the electrophysiological/biophysical differences (e.g. gram positive and gram negative) between various bacteria types, as a basis to distinguish between bacteria concentrations and bacteria types. A close-ended, coaxial probe of 20.0 mm long sample-holder was developed and used to calculate the dielectric constant from the measured S parameters of the bacterial cultures, using the Nicolson-Ross-Weir method. This technique shows a linear relationship (r2 ≥ 0.999) between the dielectric constant and the cell concentration, at 16.0 GHz. The sensitivity of the technique is 0.177 × 109 (CFU mL−1)−1 for B. Subtilis (with a size of 10.0 × 1.00 μm), 0.322 × 109 (CFU mL−1)−1 for E. Coli (with a size of 2.00 × 0.500 μm) and 0.913 × 109 (CFU mL−1) −1 for their 1:1 mixture, while the response time is 60.0s. The dependency of dielectric constant on the bacterial cell concentration at a given frequency can be potentially exploited for measuring bacterial concentrations and biophysical differences.