Method for Estimating the Acoustic Pressure in Tissues Using Low-Amplitude Measurements in Water
Date
2015Author
Keravnou, Christina P.Izamis, Maria-Louisa
AVERKIOU, MICHALAKIS A.
Source
Ultrasound in medicine & biologyVolume
41Pages
3001-3012Google Scholar check
Keyword(s):
Metadata
Show full item recordAbstract
The aim of this study was to evaluate a simple, reliable and reproducible method for accuracy in estimating the acoustic pressure delivered in tissue exposed to ultrasound. Such a method would be useful for therapeutic applications of ultrasound with microbubbles, for example, sonoporation. The method is based on (i) low-amplitude water measurements that are easily made and do not suffer from non-linear propagation effects, and (ii) the attenuation coefficient of the tissue of interest. The range of validity of the extrapolation method for different attenuation and pressure values was evaluated with a non-linear propagation theoretical model. Depending on the specific tissue attenuation, the method produces good estimates of pressures in excess of 10 MPa. Ex vivo machine-perfused pig liver tissue was used to validate the method for source pressures up to 3.5 MPa. The method can be used to estimate the delivered pressure in vivo in diagnostic and therapeutic applications of ultrasound.