Quality evaluation of ultrasound imaging in the carotid artery based on normalization and speckle reduction filtering
Date
2006Author
Loizou, Christos P.Pattichis, Constantinos S.
Pantzaris, Marios C.
Tyllis, Theodosis K.
Nicolaïdes, Andrew N.
Source
Medical and Biological Engineering and ComputingVolume
44Issue
5Pages
414-426Google Scholar check
Keyword(s):
Metadata
Show full item recordAbstract
Image quality is important when evaluating ultrasound images of the carotid for the assessment of the degree of atherosclerotic disease, or when transferring images through a telemedicine channel, and/or in other image processing tasks. The objective of this study was to investigate the usefulness of image quality evaluation based on image quality metrics and visual perception, in ultrasound imaging of the carotid artery after normalization and speckle reduction filtering. Image quality was evaluated based on statistical and texture features, image quality evaluation metrics, and visual perception evaluation made by two experts. These were computed on 80 longitudinal ultrasound images of the carotid bifurcation recorded from two different ultrasound scanners, the HDI ATL-3000 and the HDI ATL-5000 scanner, before (NF) and after (DS) speckle reduction filtering, after normalization (N), and after normalization and speckle reduction filtering (NDS). The results of this study showed that: (1) the normalized speckle reduction, NDS, images were rated visually better on both scanners (2) the NDS images showed better statistical and texture analysis results on both scanners (3) better image quality evaluation results were obtained between the original (NF) and normalized (N) images, i.e. NF-N, for both scanners, followed by the NF-DS images for the ATL HDI-5000 scanner and the NF-DS on the HDI ATL-3000 scanner (4) the ATL HDI-5000 scanner images have considerable higher entropy than the ATL HDI-3000 scanner and thus more information content. However, based on the visual evaluation by the two experts, both scanners were rated similarly. The above findings are also in agreement with the visual perception evaluation, carried out by the two vascular experts. The results of this study showed that ultrasound image normalization and speckle reduction filtering are important preprocessing steps favoring image quality, and should be further investigated. © International Federation for Medical and Biological Engineering 2006.
Collections
Cite as
Related items
Showing items related by title, author, creator and subject.
-
Article
Manual and automated intima-media thickness and diameter measurements of the common carotid artery in patients with renal failure disease
Loizou, Christos P.; Kasparis, Takis; Lazarou, Theodoros; Pattichis, Constantinos S.; Pantzaris, Marios C. (2014)The objective of this study was to investigate differences in intima-media thickness (IMT) and diameter (D) measurements of the common carotid artery (CCA) in ultrasound imaging in normal subjects and renal failure disease ...
-
Article
Impact of head rotation on the individualized common carotid flow and carotid bifurcation hemodynamics
Aristokleous, Nicolas; Seimenis, Ioannis; Georgiou, Georgios C.; Papaharilaou, Yannis; Brott, B. C.; Nicolaïdes, Andrew N.; Anayiotos, Andreas S. (2014)This paper aims at evaluating the changes that head rotation poses on morphological and flow characteristics of the carotid bifurcation (CB) and on the distribution of parameters that are regarded as important in atherosclerosis ...
-
Article
Snakes based segmentation of the common carotid artery intima media
Loizou, Christos P.; Pattichis, Constantinos S.; Pantzaris, Marios C.; Tyllis, Theodosis K.; Nicolaïdes, Andrew N. (2007)Ultrasound measurements of the human carotid artery walls are conventionally obtained by manually tracing interfaces between tissue layers. In this study we present a snakes segmentation technique for detecting the ...